Electronics Revisited, LLC

HP 8640B  Test and Verification Procedure

 

Serial Number:_______________________

Date: _____________   Tech:___________

Source:_____________________________

 

Mechanical:

____  C/L AM slider

____  C/L FM slider

____  C/L Varactor gears

____  C/L Fine tuning

____  C/L FM Range and Frequency Assembly

____  Check all contact fingers on FM Range and Freq Assemb

____  Check contact fingers on Attenuator

____  Check all gears on FM Range and Freq Assemb.

____  Check all gears on Varactor Assemb

____  C/L Modulation Frequency Selector

____  C/L Lube Mechanics on Var Mod (option 001 only)

____  C/L three meter select push buttons

____  Verify button caps remain tight on meter select

____  C/L six buttons on freq. counter assembly

____  Check for loose/missing hardware

____  Check all gearing for slippage

____  C/L top and bottom covers

____  C/L front panel

____  Check screw tightness,all flange mtd. transistors

____  Check power switch and lamp

____  Check for missing external or Internal hardware

____  Check Analog Meter "Zero" reading

 

Electrical:

____  Check all power supply voltages

____  Check all six meter select lamps

          ____ Check "Reduce Peak Power" Lamp

          ____ Check "Reduce AM Vernier" Lamp

          ____ Check "Reduce Peak Deviation" Lamp

          ____ Check all three range lamps

____  Check all 7-seg frequency display assemb

____  Verify output for all output sel. w/correct meter readings

____  Check Analog Meter smooth w/ no snags or skips.

____  Check frequency display/output for all frequency select ranges

____  Calibrate frequency against GPS Standard

          ____ Check x10 Scale select

          ____ Check x100 Scale select 

          ____ Affix calibration label (frequency only)        

____  Check frequency doubler (opt 002 only)

____  Check frequency display uncal selector

____  Check and cal int AM mod for all mod freq w/scope

        ____ Null at A2TP3 w/A2R15 per manual

        ____ Null at A2TP4 w/A2R14 per manual

        ____ Check at A26A2TP1, adj A2R29

        ____ Check/Adjust A26R19 for 100% AM mod

____  Check modulation frequency at front panel BNC

        _____ 400 Hz

        _____ 1000 Hz

        _____ All other ranges if Opt 001 installed

____  Check External AM/AC input

____  Check External AM/DC input

____  Check Pulse modulation

____  Check Internal FM modulation with Ext mod meter

____  Check External FM/AC input

____  Check External FM/DC input

____  Check FM/Cal setting

____  Check external frequency counter input functional

          ____ check 0-10 range

          ____ check 10-550 range

____  Check phase lock function

____  Check ½ digit function if so equipt

____  Check rear panel BNC Auxiliary output

____  Check rear BNC ext reference w/ Int/Ext switch

____  Check power cord for nicks, etc.

 

 

____  Perform 2-4 hour burn-in

           ____ Re-Verify freq cal against GPS reference

           ____ Re-Verify Internal FM/AM/Pulse modulation

           ____ Re-Verify all outputs

           ____ Rd-Verify all meter functions and lamps

 

Pre-Shipment Tests:

            ____ Re-verify power up and freq outputs

            ____ Apply warranty tags

 

___ Opt 001              _________________

___ Opt 002              _________ RRA #

___ Opt 003              ________________ Client

___ Opt 004

 

Definitions:

 

C/L = C/L and Lube

Check = Check, adjust, replace and/or repair as required.

Verify= Confirm Functional, repair/replace as required. (May also require Re-check if post burn-in)

 

Notes: